Título del libro: 2014 Ieee 9th Iberoamerican Congress On Sensors, Ibersensor 2014 - Conference Proceedings
Título del capítulo: Design of an angled ultrasonic transducer for blood flow measurement applications

Autores UNAM:
PEDRO JESUS ACEVEDO CONTLA; ISRAEL SANCHEZ DOMINGUEZ;
Autores externos:

Idioma:
Inglés
Año de publicación:
2014
Palabras clave:

Blood; Design; Hemodynamics; Piezoelectric ceramics; Transducers; Ultrasonic transducers; Blood flow measurement; Coronary bypass; Design and construction; Matching layer; Pulsed-wave Doppler; rexolite; Sensing elements; Single element transducers; Flow measurement


Resumen:

This paper describes the design and construction of an angled ultrasonic transducer for blood flow measurement. The transducer was constructed using as sensing element, a commercial 8 MHz PIC255 piezoceramic, Insulcast 501 and Rexolite 4422 as backing and matching layer materials respectively. The objective of this work was to design and construct a single element transducer for pulsed wave Doppler blood flow measurement, with an angle of 60° resonating at approximately 8 MHz. Results show that the overall performance of the transducer fulfils the requirements of its design and it is possible to use it for blood flow measurement applications. © 2014 IEEE.